Title: | Uncertainty evaluation of proposed setup for the calibration of vertical angle measuring systems by using means for the flat angle calibration |
Authors: | Šiaudinytė, Lauryna Suh, Ho Suhng |
Keywords: | Autocollimator Calibration Measurement Total station Precision index Vertical angle |
Issue Date: | 2015 |
Publisher: | Elsevier Ltd |
Citation: | 10. Siaudinyte L., Suh H.S. 2015. Uncertainty evaluation of proposed setup for the calibration of vertical angle measuring systems by using means for the flat angle calibration, MEASUREMENT 67: 177–182. DOI: 10.1016/j.measurement.2015.01.001 . |
Series/Report no.: | 67; |
Abstract: | Precision angle measurements performed by using total stations is a field showing a lot of
potential. Horizontal angle measuring system calibration is a well documented and
researched field however vertical angle measuring system calibration tends to lack proper
investigation. The paper deals with the analysis of modern calibration methods of vertical
angle measuring systems. New setup for the calibration of vertical angle measuring system
of the total station using horizontal angle calibration means is presented in the paper. The
principle of the method is explained and the uncertainty evaluation as well as preliminary
results are given in the paper. The advantages and weaknesses of this new setup are
discussed. |