VGTU talpykla > Aplinkos inžinerijos fakultetas / Faculty of Environmental Engineering > Moksliniai straipsniai / Research articles >

Lietuvių   English
Please use this identifier to cite or link to this item: http://dspace.vgtu.lt/handle/1/3781

Title: Uncertainty evaluation of proposed setup for the calibration of vertical angle measuring systems by using means for the flat angle calibration
Authors: Šiaudinytė, Lauryna
Suh, Ho Suhng
Keywords: Autocollimator
Calibration
Measurement
Total station
Precision index
Vertical angle
Issue Date: 2015
Publisher: Elsevier Ltd
Citation: 10. Siaudinyte L., Suh H.S. 2015. Uncertainty evaluation of proposed setup for the calibration of vertical angle measuring systems by using means for the flat angle calibration, MEASUREMENT 67: 177–182. DOI: 10.1016/j.measurement.2015.01.001 .
Series/Report no.: 67;
Abstract: Precision angle measurements performed by using total stations is a field showing a lot of potential. Horizontal angle measuring system calibration is a well documented and researched field however vertical angle measuring system calibration tends to lack proper investigation. The paper deals with the analysis of modern calibration methods of vertical angle measuring systems. New setup for the calibration of vertical angle measuring system of the total station using horizontal angle calibration means is presented in the paper. The principle of the method is explained and the uncertainty evaluation as well as preliminary results are given in the paper. The advantages and weaknesses of this new setup are discussed.
URI: http://dspace.vgtu.lt/handle/1/3781
ISSN: 0263-2241
Appears in Collections:Moksliniai straipsniai / Research articles

Files in This Item:

File Description SizeFormat
1-s2.0-S0263224115000196-main.pdf718.86 kBAdobe PDFView/Open

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

 